Model 2651A High Power System SourceMeter® Instrument Reference Manual
The safety shield can be metallic or nonmetallic, and must completely surround the DUT test circuit. A
metal safety shield must be connected to a known safety earth ground and chassis ground. See
fixture
(on page 2-57) later in this section for important safety information on the use of a metal or
nonmetallic enclosure.
Model 2651A safety shield
The maximum output voltage for a Model 2651A channel is 40 V, which is considered a
nonhazardous level. However, using two Model 2651A voltage sources in a series configuration or
floating a SMU
the source-measure units (SMUs) of two Model 2651A instruments can be connected in series to
apply 80 V to a device under test (DUT) (see
on using multiple Model 2651A).
Use #18 AWG wire or larger for connections to safety earth ground and chassis.
Figure 21: Safety shield for hazardous voltage using two Model 2651A instruments
2651A-901-01 Rev. A / March 2011
(on page 2-58) can cause test circuit voltage to exceed 42 V. In the following figure,
Model 2651A (master)
Chassis
ground
SENSE/GUARD
S
LO
Chassis
(subordinate)
Chassis
earth
ground
TSP advanced features
Model 2651A (subordinate)
SENSE/GUARD
S
LO
G
G G G
OUTPUT
S
G
G G G
HI
40V, 50A MAX.
LO
Equivalent
HI
SMU
40 V
HI
LO
80 V
DUT
LO
LO
SMU
40 V
(master)
HI
Section 2: General operation
(on page 6-53) for information
OUTPUT
LO
HI
S
HI
40V, 50A MAX.
LO
LO
HI
HI
LO
HI
DUT
Safety
earth
ground
Metal safety shield
Safety
earth
ground
Test
HI
2-55