Section 7: Command reference
Details
This attribute can be set to initialize the pseudo state to a known value.
Setting this attribute will not cause the LAN trigger to generate any events or output packets.
Example
print(lan.trigger[1].pseudostate)
Also see
None
lan.trigger[N].stimulus
This attribute specifies events that cause this trigger to assert.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
triggerStimulus = lan.trigger[N].stimulus
lan.trigger[N].stimulus = triggerStimulus
triggerStimulus
N
Details
This attribute specifies which event causes a LAN trigger packet to be sent for this trigger. Set
triggerStimulus to one of the existing trigger event IDs shown in the following table.
Trigger event IDs*
Event ID
smuX.trigger.SWEEPING_EVENT_ID
smuX.trigger.ARMED_EVENT_ID
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
7-130
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Get the present simulated line state for the LAN
event 1.
Affected by
Instrument reset
LAN trigger N reset
The LAN event identifier used to trigger the event
A number specifying the trigger packet over the LAN for which to set or query
the trigger source (1 to 8)
Where saved
Default value
Not saved
0
Event description
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
Occurs when the SMU completes a source action
Occurs when the SMU completes a measure action
2651A-901-01 Rev. A / March 2011