Section 2: General operation
Source range commands*
Commands
smua.source.autorangei = smua.AUTORANGE_ON
smua.source.autorangei = smua.AUTORANGE_OFF
smua.source.autorangev = smua.AUTORANGE_ON
smua.source.autorangev = smua.AUTORANGE_OFF
smua.source.limiti = level
smua.source.limitv = level
smua.source.lowrangei = lowrange
smua.source.lowrangev = lowrange
smua.source.rangei = rangeval
smua.source.rangev = rangeval
* See
Available ranges
Range programming example
The programming example below illustrates how to control both source and measure ranges. The
Model 2651A is set up as follows:
Voltage source range: Auto
Current measure range: 10 mA
Voltage source current limit: 10 mA
-- Restore Model 2651A defaults.
smua.reset()
-- Set V source range to auto.
smua.source.autorangev = smua.AUTORANGE_ON
-- Select 10 mA measure range.
smua.measure.rangei = 1e-2
-- Set limit level to 10 mA.
smua.source.limiti = 1e-2
2-70
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 2-65).
Description
Enable current source
autorange.
Disable current source
autorange.
Enable voltage source
autorange.
Disable voltage source
autorange.
Set voltage source current
limit.
Set current source voltage
limit.
Set lowest I source range for
autorange.
Set lowest V source range for
autorange.
Select manual current source
range.
Select manual voltage source
range.