Advertisement

Section 7: Command reference
Usage
SweepIListMeasureV(smu, ilist, stime, points)
smu
ilist
stime
points
Details
Data for voltage measurements, current source values, and timestamps are stored in smuX.nvbuffer1.
If all parameters are omitted when this function is called, this function is executed with the parameters set to the
default values.
Performs a current list sweep with voltage measured at every step (point):
1. Sets the smu to output ilist amperes value, allows the source to settle for stime seconds, and
then performs a voltage measurement.
2. Sets the smu to output the next ilist step, allows the source to settle for stime seconds, and
then performs a voltage measurement.
3. Repeats the above sequence until the voltage is measured for the last amperes value. The last
point in the list to be measured is points.
Example
myilist = {-100E-9, 100E-9, -1E-6, 1E-6,
-1E-3, 1E-3}
SweepIListMeasureV(smua, myilist,
500E-3, 6)
Also see
None

SweepILogMeasureV()

This
KISweep factory script
at every step (point).
Type
TSP-Link accessible
Function
No
7-326
Model 2651A High Power System SourceMeter® Instrument Reference Manual
System SourceMeter
Arbitrary list of current source values; ilist =
....valueN}
Settling time in seconds; occurs after stepping the source and before performing a
measurement
Number of sweep points (must be ≥2)
(on page 5-22) function performs a logarithmic current sweep with voltage measured
Affected by
®
instrument channel (set to smua)
This function performs a six-point current list sweep
starting at the first point in myilist. Voltage is
measured at every step (point) in the sweep. The
source will be allowed to settle on each step for
500 ms before a measurement is performed.
Where saved
{value1, value2,
Default value
2651A-901-01 Rev. A / March 2011