Enabling High-Capacitance Mode - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 3: Functions and features

Enabling high-capacitance mode

Before enabling high-capacitance mode, note the following:
It is important to read the previous section to understand the impact of high-capacitance mode.
Test the DUT and the capacitor to determine the best current source and range of output voltages.
The settling times can vary based on the DUT. It is important to test the limits of the DUT before you
use high-capacitance mode.
Failure to test the DUT for the appropriate current source and output voltages can result in damage to or
destruction of the DUT.
For optimal performance, do not continuously switch between normal mode and high-capacitance
mode.
Before you charge the capacitor, start with 0 (zero) voltage across the capacitor.
When high-capacitance mode is enabled, a minimum load capacitance of 100 nF is recommended. In
absence of this minimum load capacitance, overshoot and/or ringing may occur.
Front panel
To enable high-capacitance mode from the front panel:
1. Press the CONFIG key, then select SRC > HIGHC-MODE.
2. Select SRC-ENABLE > ENABLE. High-capacitance mode is enabled.
3. Push the ENTER key.
4. Press the EXIT (LOCAL) key to back out of the menu structure.
Command interface
Turning on High-C mode has the following effects on the SMU settings:
smua.measure.autorangei is set to smua.AUTORANGE_FOLLOW_LIMIT and cannot be
changed.
Current ranges below 1 μA are not accessible.
If smua.source.limiti is less than 1 μA, it is raised to 1 μA.
If smua.source.rangei is less than 1 μA, it is raised to 1 μA.
If smua.source.lowrangei is less than 1 μA, it is raised to 1 μA.
If smua.measure.lowrangei is less than 1 μA, it is raised to 1 μA.
3-68
Model 2651A High Power System SourceMeter® Instrument Reference Manual
2651A-901-01 Rev. A / March 2011

Advertisement

Table of Contents
loading

Table of Contents