smuX.trigger.measure.action
This attribute controls measurement actions during a sweep.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
action = smuX.trigger.measure.action
smuX.trigger.measure.action = action
action
X
2651A-901-01 Rev. A / March 2011
(on page 7-20)
(on page 3-29)
(on page 7-235)
(on page 7-238)
(on page 7-241)
(on page 7-243)
(on page 7-244)
(on page 7-245)
Affected by
SMU reset
Instrument reset
The sweep measure action; set to one of the following values:
0 or smuX.DISABLE: Do not make measurements during the sweep
1 or smuX.ENABLE: Make measurements during the sweep
2 or smuX.ASYNC: Make measurements during the sweep, but asynchronously
with the source area of the trigger model
Source-measure unit (SMU) channel (for example,
smua.trigger.measure.action applies to SMU channel A)
Where saved
Default value
Not saved
smuX.DISABLE
Section 7: Command reference
7-235