Smux.source.func - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Usage
sDelay = smuX.source.delay
smuX.source.delay = sDelay
sDelay
X
Details
This attribute allows for additional delay (settling time) after an output step. Set sDelay to a user-defined value
(in seconds). Alternatively, set sDelay to smuX.DELAY_OFF or smuX.DELAY_AUTO.
The smuX.DELAY_AUTO setting causes a range-dependent delay to be inserted when the source is changed.
Range dependent delays are based on the output settling time values in the instrument's specifications.
Example
smua.source.delay = smua.DELAY_AUTO
Also see
reset()
(on page 7-151)
smuX.measure.count
smuX.measure.delay
smuX.reset()

smuX.source.func

This attribute sets the source function (V source or I source).
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sFunction = smuX.source.func
smuX.source.func = sFunction
sFunction
X
7-214
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Set to your desired source delay value. For example, to specify an additional 10 ms
source delay, set the value to 0.010.
Alternatively, you can set it one of the following values:
0 or smuX.DELAY_OFF: No delay
-1 or smuX.DELAY_AUTO: Automatic delay value
SMU channel (for example, smua.source.delay applies to SMU Channel A).
(on page 7-194)
(on page 7-195)
(on page 7-209)
Affected by
SMU reset
Instrument reset
Recall
The source function. Set to one of the following values:
0 or smuX.OUTPUT_DCAMPS: Selects current source function.
1 or smuX.OUTPUT_DCVOLTS: Selects voltage source function.
SMU channel (for example, smua.source.func applies to SMU Channel A).
Sets the delay for SMU Channel A
to automatic (a range dependent
delay is inserted when ever the
source is changed).
Where saved
Default value
Saved setups
smuX.OUTPUT_DCVOLTS
2651A-901-01 Rev. A / March 2011