Smux.reset(); Smux.savebuffer() - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual

smuX.reset()

This function turns off the output and resets the SMU to the default settings.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.reset()
X
Details
This function turns off the output and returns the specified SMU to its default settings.
Example
smua.reset()
Also see
reset()
(on page 7-151)

smuX.savebuffer()

This function saves one source-measure unit (SMU) dedicated reading buffer to internal memory (there are two
dedicated reading buffers per SMU).
Type
TSP-Link accessible
Function
Yes
Usage
smuX.savebuffer(smuX.nvbufferY)
X
Y
Details
When the unit is turned off and back on, the dedicated reading buffers will be restored (from internal memory) to
their last saved values.
Example
smua.savebuffer(smua.nvbuffer1)
2651A-901-01 Rev. A / March 2011
Affected by
Source-measure unit (SMU) channel (for example, smua.reset() applies to SMU
channel A).
Affected by
SMU channel (for example, smua.savebuffer(smua.nvbuffer1) applies to
SMU channel A)
SMU dedicated reading buffer (1 or 2)
Where saved
Default value
Turns off the output and resets SMU
channel A to its default settings.
Where saved
Default value
Saves buffer 1 (SMU channel A) to
internal memory.