Voltage source accuracy

Follow the steps below to verify that the Model 2651A output voltage accuracy is within specified
limits. This test involves setting the output voltage to each full-range value and measuring the
voltages with a precision digital multimeter.
1. With the power off, connect the digital multimeter (DMM) to the Model 2651A output terminals
using 4-wire connections, as shown below.
2. Select the multimeter DC volts measuring function.
3. Press the SRC key to source voltage and make sure the source output is turned on.
4. Enable the Model 2651A 4-wire (remote sense) mode by pressing the CONFIG key and then the
SRC key.
5. Select V-SOURCE > SENSE-MODE > 4-WIRE.
2651A-901-01 Rev. A / March 2011
Figure 114: Connections for voltage verification