Keithley 2651A Reference Manual page 496

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
smuX.contact.speed
This attribute stores the speed setting for contact check measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
speedSetting = smuX.contact.speed
smuX.contact.speed = speedSetting
speedSetting
X
Details
This setting controls the aperture of measurements made for contact check. It does not affect the
smuX.measure.nplc aperture setting.
The speed setting can have a dramatic effect on the accuracy of the measurement (see specifications).
Example
smua.contact.speed = smua.CONTACT_SLOW
Also see
reset()
(on page 7-151)
smuX.contact.check()
smuX.contact.r()
smuX.reset()
Contact check connections
Contact check measurements
7-188
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Instrument reset
SMU reset
The speed setting. Set to one of the following:
0 or smuX.CONTACT_FAST
1 or smuX.CONTACT_MEDIUM
2 or smuX.CONTACT_SLOW
SMU channel (for example, smua.contact.speed applies to SMU Channel A).
(on page 7-186)
(on page 7-187)
(on page 7-209)
(on page 2-45)
(on page 2-37)
Where saved
Default value
Not saved
smuX.CONTACT_FAST
Configure contact check for higher accuracy
on SMU Channel A.
2651A-901-01 Rev. A / March 2011

Advertisement

Table of Contents
loading

Table of Contents