Section 7: Command reference
The SMU will only store the most recent configured source action. The last call to
smuX.trigger.source.linearY(), smuX.trigger.source.listY(), or
smuX.trigger.source.logY() is used for the source action.
Source functions cannot be changed within a sweep.
After configuring the sweep source values, enable the source action by setting
smuX.trigger.source.action.
Example
smua.trigger.source.logv(1, 10, 11, 0)
Also see
smuX.trigger.source.action
smuX.trigger.source.linearY()
smuX.trigger.source.listY()
Sweep operation

smuX.trigger.source.set()

This function sets the source event detector to the detected state.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.trigger.source.set()
X
Details
This function sets the source event detector to the detected state.
The SMU will automatically clear all event detectors when the smuX.trigger.initiate() function is
executed. This function should be called after the sweep is initiated. Make sure that if the event detectors are
configured to automatically clear (because the smuX.trigger.autoclear attribute is set to smuX.ENABLE)
that this command is issued after the SMU has entered the trigger layer.
Also see
smuX.trigger.arm.set()
smuX.trigger.autoclear
smuX.trigger.endpulse.set()
smuX.trigger.initiate()
smuX.trigger.measure.set()
Triggering
(on page 3-32)
7-246
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 7-241)
(on page 7-243)
(on page 7-244)
(on page 3-19)
Affected by
Source-measure unit (SMU) channel (for example,
smua.trigger.source.set() applies to SMU channel A)
(on page 7-226)
(on page 7-229)
(on page 7-231)
(on page 7-234)
(on page 7-236)
Sweeps SMU channel A from 1 V
to 10 V in 10 steps with an
asymptote of 0 V.
Where saved
Default value
2651A-901-01 Rev. A / March 2011