Cleaning The Skimmer And Tube Lens On The Lxq And Ltq Xl Mass Spectrometers; Figure 54. Ion Source Interface Removed From The Vacuum Manifold - Thermo Scientific LTQ Series Hardware Manual

Table of Contents

Advertisement

5
Maintenance
Ion Source Interface Maintenance

Cleaning the Skimmer and Tube Lens on the LXQ and LTQ XL Mass Spectrometers

80
LTQ Series Hardware Manual
4. Grasp the ridges on either side of the ion source interface with your fingers and carefully
pull it out of the vacuum manifold

Figure 54. Ion source interface removed from the vacuum manifold

Ion source interface
(inside cage)
5. Place the ion source interface on the clean, lint-free work surface.
Chemicals accumulating on the surfaces of the skimmer and tube lens form an insulating layer
that can modify the electrical fields controlling ion transmission. The skimmer and tube lens
require cleaning less often than the ion sweep cone and the ion transfer tube. No tools are
needed to remove or install these components.
CAUTION Do not scratch the outside of the skimmer cone.
To clean the skimmer and tube lens, follow these procedures:
1.
To remove and clean the tube lens and skimmer
2.
To reinstall the tube lens and
(Figure
54).
Vacuum
manifold
skimmer, on
page 83
Outer cage (installed)
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents