3. 3. 12. Gassing Manifold Installation
The gassing manifold allows for injecting an inert gas such as nitrogen or
carbon dioxide into multiple vessels during shaker use (see chapter
General Accessories" on page 42
The gassing manifold is a factory-installed option for Solaris 2000 I/R and
4000 I/R shakers that is also available as a customer-installed accessory
when not originally installed on the unit. All Solaris 2000 I/R and 4000 I/R
shakers have an access port on either side panel (see chapter
Overview" on page 44
gas supply tube to the manifold assembly. The gassing manifold has one
inlet fitting for the gas supply and eight outlet fittings for vessels.
Figure 100: Gassing Manifold Installed in Solaris 4000 R Shaker
Install as follows:
1.
Securely fasten the gas manifold ➁ to the rear wall of the shaker's
chamber with the screws ➄ supplied.
2.
Remove the plastic plug from the outer side of the access port grommet
(see chapter
information).
3.
Make a crossed incision with a cutter knife into the membrane of the
rubber grommet inside the shaker chamber.
4.
Cut a piece of flexible tube to a suitable length for use as the lab gas
supply tube ➀.
Accessories
for ordering details).
for location information) for feeding through the lab
"1. 4. Product Overview" on page 44
115 / 143
"1. 2. 12.
"1. 4. Product
➀  Lab gas supply tube
➁  Gassing manifold assy
➂  Outlet fittings with tubes
➃  Inlet fitting
➄  Fastening screws
for location