Lens Driver Board - Thermo Scientific TSQ 8000 Evo Hardware Manual

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Boards in the TSQ 8000 Evo Instrument

Lens Driver Board

282
TSQ 8000 Evo Hardware Manual
To replace the lens driver board
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107. You do not have to remove the ion source cartridge or the column.
2. Slide the top cover panel toward the back of the instrument and it them off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 302. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
"Shutting Down the TSQ 8000 Evo
Screws
Screw on Front
of Instrument