Removing the MP00 RF Lens Assembly

Thermo Scientific
For Field Service use only
O-ring, Viton, 3.25 in. ID, 1/8 in. thick
To remove the MP00 rf lens assembly
CAUTION To avoid an electrical shock, be sure to follow the instructions in
Down the Mass Spectrometer Completely"
procedure.
1. Shut down and vent the system, and let it cool to room temperature.
Venting the mass spectrometer can take several minutes.
2. Unplug the mass spectrometer's power supply cord from the electrical outlet.
3. Remove the ion source interface (see
4. Reach into the opening in the vacuum manifold (where the ion source interface was) and
disconnect the electrical connector to the MP00 rf lens assembly by pulling it toward the
left.
5. Using the 5/32 in. hex ball driver, loosen the three hex socket-head screws that secure the
ion source interface to the vacuum manifold's outer cage
on
page 36
before continuing with this
page
79).
(Figure
LTQ Series Hardware Manual
5
Maintenance
MP00 RF Lens Maintenance
Part number
00107-13000
"Shutting
64).
93