Preventive Maintenance and Servicing
Maintenance and Servicing of the FTIR Analyzer Components
Replacing the
5-16
Omni FTIR Multi Gas CEMS Instruction Manual
for damage to sensitive components. Damage to components inside the
analyzer will affect the performance of the system.
Use the following procedure to replace the source in the FTIR analyzer.
The source is located behind the right side panel of the analyzer.
Source
WARNING To avoid shock hazard, always turn off the FTIR circuit
breaker on the breaker panel before replacing any components inside the
FTIR analyzer.
CAUTION The source element becomes extremely hot during normal
analyzer operation. Always turn off the analyzer and allow the elements to
cool for at least 15 minutes before removing the source from the analyzer.
Handle the source assembly only by the ceramic base.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal components.
Important Note The FTIR analyzer source is pinned in place and pre-
aligned. Replacing the source should not affect measurement results,
however, an improperly installed or handled source will affect the
performance of the system. A system operator can replace the source,
however, it is recommended that only our certified service engineers or on-
site maintenance personnel who have been trained by us replace the
source.
Important Note Never touch the new source element with your bare
fingers. Skin oils or other deposits on the element will shorten its life.
Equipment Required:
Phillips screwdriver
Hex wrench, 3/32-inch
Source assembly
1. Using the Phillips screwdriver, remove the FTIR cover panel in the
front of the system rack. Slide the FTIR drawer out of the rack to gain
access to the analyzer.