Thermo Scientific
To clean the spray cone and O-ring
IMPORTANT
Always use LC/MS-grade methanol and LC/MS-grade water (see
Table 7
on
page
55).
1. Soak Kimwipe tissues or chamois-tipped swabs in a 50:50 solution of methanol/water,
and then clean the exterior surface of the spray cone.
2. Remove and inspect the O-ring that is in the spray cone under the entrance end of the ion
transfer tube
(Figure 62
Figure 62. Ion source interface components for the LXQ and LTQ XL
Ion transfer tube
Figure 63. Ion source interface components for the Velos Pro
Ion transfer tube
3. Clean the O-ring with methanol or replace it if necessary (see
and
Figure
63).
5
Maintenance
Ion Source Interface Maintenance
Spray cone
Vespel O-ring
Ion sweep cone
Spray cone
Vespel O-ring
Ion sweep cone
page
85).
LTQ Series Hardware Manual
89