2. To turn the power off, press the main power switch to position 0 (Off).
None of the front-panel LEDs should be on.
2.1.6

Function Check

The instrument automatically monitors the main functions when it is switched on
and monitors them continuously during operation.
A detected fault is indicated by an "Error" message displayed in the info line of
the R&S SGMA-GUI together with a brief error description. For an in-depth identi-
fication of the error, press the "SGMA-GUI > Info" button. In response, a descrip-
tion of the errors is displayed. For more information, refer to the "Error Messages"
section in the user manual.
In addition to the automatic monitoring, the R&S SGS offers the following capabil-
ities to ensure correct functioning:
● Internal Adjustments
In the R&S SGMA-GUI, select the "Instrument > Setup > Internal Adjust-
ments" dialog to access the dialog for performing and configuring the adjust-
ments settings. A maximum level accuracy can be obtained, for instance.
● Selftest
A selftest is provided for service purposes ("SGMA-GUI > Instrument > Diag-
nostic/Test > Self Test").
2.1.7

Default Settings

When the instrument is switched on, it is not the preset state that is active, but
rather the instrument state that was set before the instrument was switched off. It
is recommended that you use the "SGMA-GUI > Instrument > Preset" function to
return the instrument to its defined preset state every time a new configuration is
required or the current setup is no longer relevant.
The R&S SGS offers a two-stage preset concept:
● Preset the instrument to a predefined state
The "SGMA-GUI > Instrument Name > Preset" function calls up a defined
instrument setup. All parameters and switching states are preset (also these
of inactive operating modes). The default instrument settings provide a repro-
ducible initial basis for all other settings. However, functions that concern the
integration of the instrument into a measurement setup are not changed.