When required
Tools required
Parts required
The interface board is sensitive to electrostatic discharge. Always use the ESD strap
CAU TI O N
when handling electronic boards.
1 Switch off the sampler at the main power switch.
2 Disconnect all cables from the existing interface board. Then loosen the
3 Identify the interface board slot cover. Loosen the two holding screws, and
4 Carefully insert the new interface board into the holding rails, and push the
5 Reconnect all cables to the new interface board.
6 Switch on the sampler.
1100 Series WS MWS Reference Manual
For all repairs inside the sampler or for installation of the board.
Flat-head screwdriver.
Interface board, see
interface board holding screws and slide the board out of its holding rails.
remove the cover.
board into the slot. Make sure the board plugs into the socket correctly.
Repairing the Sampler
"Optional Interface Boards"
on page 225