7
Error Information
Pump Error Messages

Overcurrent of solvent selection valve (SSV)

Error ID: 22412
Overcurrent of solvent selection valve (SSV).
• Parameter: None
Probable cause
1

Deliver underrun

Error ID: 22413
Internal error.
• Parameter: None
Probable cause
1

Defect connection between main board and pump drive encoder

Error ID: 22414
Defect connection between main board and pump drive encoder.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
2
92
Valve defect.
Internal error.
Defect connection between main board and
pump drive encoder.
Defect encoder.
Suggested actions
Replace the solvent selection valve.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.