To Remove The Detector Top Cover - Agilent Technologies 8860 Maintaining

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

2
Removing Covers

To Remove the Detector Top Cover

To Remove the Detector Top Cover
This cover protects the detectors, valve box, and valve assembly. To remove the detector top
cover:
1 Turn off the GC and unplug the power cord.
2 Remove the detector cover by raising the cover up, then firmly lift up on the right side of the
3 When finished with the maintenance procedure, place the detector cover vertically on the
Be careful! With the power switch on, potentially dangerous voltages also exist:
WAR N IN G
• All electronics boards in the instrument.
• The internal wires and cables connected to these boards.
• The wires for the oven heater.
The covers shield these parts as the parts may be hot enough to cause burns. Never remove
a cover unless the instrument is unplugged.
Do not force the cover, either when installing it or closing it. This could break the plastic
C AU T I O N
parts.
Pneumatics cover
Injection
port cover
Figure 2.
18
cover to free it from the GC. Set the cover aside.
GC and insert the left side first, then gently snap in the right side onto the GC.
8860 GC covers
Detector cover
Electronics cover
Maintaining Your GC