9529 Background Lamp interlock

Instrument Fault. The instrument detected a problem with the
Background Corrector (D
that the D
will not light in this position. Remove and re-install the D
bracket as described in Chapter 6. If the problem persists a service
call may be required.

9530 Instrument Fault: Signal diagnostics

Instrument Fault. The instrument detected a problem with the signal
electronics on the control board. Restart the instrument. If the
problem persists a service call will be required.

9531 No Lamp Current detected

The instrument could not detect any lamp current for the selected
Hollow Cathode lamp and has aborted the peaking. Rectify the
problem and optimize the instrument.
Possible causes include:

9602 Instrument Fault: Mono Datum

Instrument Fault. The instrument detected a problem with the
monochromator. Restart the instrument. If the problem persists a
service call will be required.

9911 No Burner Fitted

An attempt was made to ignite the flame without any burner
installed in the sample compartment.
Agilent 55B AA Spectrometer User's Guide
lamp is not safely installed in the instrument. The lamp
2
No hollow cathode lamp in the selected lamp position
Faulty hollow cathode lamp.
) lamp. The safety interlock switch suggests
2
Troubleshooting/Errors
lamp
2
133