31
:TRIGger Commands

:TRIGger:PATTern Commands

Table 131 :TRIGger:PATTern Commands Summary
Command
:TRIGger:PATTern
<string>[,<edge_sourc
e>,<edge>] (see
page
919)
:TRIGger:PATTern:FORM
at <base> (see
page
921)
:TRIGger:PATTern:GREa
terthan
<greater_than_time>[s
uffix] (see
page
:TRIGger:PATTern:LESS
than
<less_than_time>[suff
ix] (see
page
923)
:TRIGger:PATTern:QUAL
ifier <qualifier>
(see
page
924)
:TRIGger:PATTern:RANG
e
<less_than_time>[suff
ix],
<greater_than_time>[s
uffix] (see
page
918
Query
:TRIGger:PATTern?
(see
page
920)
:TRIGger:PATTern:FORM
at? (see
page
:TRIGger:PATTern:GREa
terthan? (see
page
922)
922)
:TRIGger:PATTern:LESS
than? (see
page
:TRIGger:PATTern:QUAL
ifier? (see
:TRIGger:PATTern:RANG
e? (see
page
926)
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
Options and Query Returns
<string> ::= "nn...n" where n ::=
{0 | 1 | X | R | F} when <base> =
ASCii <string> ::= "0xnn...n"
where n ::= {0,..,9 | A,..,F | X
| $} when <base> = HEX
<edge_source> ::= {CHANnel<n> |
NONE} for DSO models
<edge_source> ::= {CHANnel<n> |
DIGital<d> | NONE} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<edge> ::= {POSitive | NEGative}
<base> ::= {ASCii | HEX}
921)
<greater_than_time> ::=
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
<less_than_time> ::=
923)
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
<qualifier> ::= {ENTered |
page
924)
GREaterthan | LESSthan | INRange
| OUTRange | TIMeout}
<less_than_time> ::= 15 ns to
926)
10 seconds in NR3 format
<greater_than_time> ::= 10 ns to
9.99 seconds in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}