:TRIGger Commands

:TRIGger:EBURst:IDLE

Command Syntax
:TRIGger:EBURst:IDLE <time_value>
<time_value> ::= time in seconds in NR3 format
The :TRIGger:EBURst:IDLE command sets the Nth edge in a burst idle
resource in seconds from 10 ns to 10 s. The timer is used to set the
minimum time before the next burst.
Query Syntax
:TRIGger:EBURst:IDLE?
The :TRIGger:EBURst:IDLE? query returns current Nth edge in a burst idle
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format