Change Pressure Measurement Tube - Weinmann homecare SOMNOvent ST Description Of The Equipment And Instructions For Use

Table of Contents

Advertisement

Fine filter /
Finfiltret /
Finfiltret
72
Coarse dust filter/
Grovdammfiltret /
Grovstøvfilter
The fine filter must be changed when it has
turned dark grey in colour, however not later
than 250 operating hours. In this case the
filter change indicator
4. Insert the filters and filter compartment
cover in reverse order.
5. To cancel the filter change indicator hold
down the on / off switch
switching on the unit until the indicator is
extinguished.
If the fine filter is changed before the 250
operating hours have elapsed due to soiling,
the hours meter has to be set to zero:
6. To do so hold down the on / off
switch
when switching on the unit.
27
The filter change indicator lights up after
3 approx. seconds and is extinguished
after another 3 seconds. Only then re-
lease the switch.
Bacteria filter
When changing the bacteria filter, please ob-
serve the relevant operating instructions.
9.3 Change pressure
measurement tube
1. Detach the sleeve on the creased hose
from the adapter
2. Remove the pressure measurement tube
from the creased hose.
11
3. Remove the pressure measurement tube
from the adapter
English
will appear.
when
27
.
10
.
10