10.3 Faults

10.3.1 Device
Fault
Alarm output too quiet
No acoustic alarm output
Alarm light does not light up
Display too dark
Device cannot be switched on
Device cannot be switched off
Red cross in function check status
report
Software update is not functioning Update file or SD card defective
Battery status indicator flickers
between red and green
Battery status indicator and the
line power indicator are not lit up
If you are not able to clear an error message with the aid of the
table, you should contact the manufacturer WEINMANN
Emergency or your authorized dealer to have the device repaired.
To avoid serious damage, do not continue using the device.
Cause
Volume set to 50%
NVG mode activated
Brightness of the display set too
low
NVG mode activated
Battery not correctly inserted in
device, or battery empty
Battery empty and device not
connected to the line power
Device defective
Operating error
Non-functioning component
Battery deeply discharged
NVG mode activated
10 Alarms and error messages
Remedy
Set the volume to 100% in the
operator menu
Deactivate NVG
p. 114).
Increase brightness of the display
in the operator menu
p. 125).
Adjust NVG brightness
p. 125).
Deactivate NVG
p. 114).
Check battery.
Check power supply.
Have the device repaired.
Press and hold the On/Off button
for at least 2 seconds.
See "9.4 Failed function check",
page 191.
Perform software update with
another SD card.
If the update still cannot be
performed successfully, have the
device repaired.
Charge battery in the device for
24 hours
(see 4.3.2,
Deactivate NVG
p. 114).
MEDUMAT Standard
(see 6.3.7,
p. 125).
(see 5.3.8,
(see 6.3.7,
(see 6.3.7,
(see 5.3.8,
p. 47).
(see 5.3.8,
2
EN
201