Function Check - Weinmann MEDUVENT Standard Step-By-Step Instructions

Hide thumbs Also See for MEDUVENT Standard:

Advertisement

Function Check

Hazardous therapy if device connected to patient during function check!
If the device is connected to the patient during the function check, this might lead to
therapy posing a risk as a result of excessively high pressures or unsuitable ventilation
volumes. This might expose the patient to the risk of serious or life-threatening injury.
Always disconnect the device from the patient for the function check.
Perform a function check at the following intervals:
Part
Device
Breathing circuit
• Before each use
• After each hygienic reprocessing
• After each repair
• Before each use
• After each hygienic reprocessing
• After each repair
Interval