Advanced Troubleshooting
Replacing Components of the Manifold

Vent Valve Knob and O-Ring

234
TSQ 8000 Evo Hardware Manual
To replace the vent valve knob and o-ring
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering these
components.
1. Shut down your system by following the steps in
System"
on
page
107. You do not have to remove the ion source cartridge or column. If
you are using hydrogen as a carrier gas, power off the GC.
.
Figure 253. Powering Off the TSQ 8000 Evo Instrument
Top Cover Panel
Screw
2. Remove and replace the vent valve knob and o-ring. Be careful not to pinch the o-ring
during installation.
3. Restart your system by following the instructions in
"Shutting Down the TSQ 8000 Evo
Vacuum Interlock Shield
Vent Valve Knob
"Restarting the System"
Power Switch
Hydrogen Safety Screws
on
page
120.