To replace the quad 1 or quad 3 wire sets
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 8000 Evo Instrument"
3. For both quad assemblies, use a T10 Torxhead screwdriver to remove the four screws
around the rods. These screws hold the wires in place. If the screws won't come off the
rods, you must replace the whole quad assembly.
4. Use on the screws provided. These screws are made from a special material designed not
fuse with the quadrupole rods.
Figure 112. Replacing the Q1 Quad Assembly Wires
Screws
"Shutting Down the TSQ 8000 Evo
on
page
Screws
109.