Ion Guide Wire Set

Thermo Scientific
To replace the ion guide wires
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering these
components.
1. Shut down your system by following the steps in
System"
on
page
107.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 8000 Evo Instrument"
3. Use a T10 Torxhead screwdriver to remove the screw connecting the ion guide wire to the
top of the ion guide.
Figure 106. Replacing the Ion Guide Wires
4. Use a wrench to remove the nut connecting the other ion guide wire to the side of the ion
guide.
5. Attach new ion guide wires and reattach the nuts and screw.
6. Follow the instructions in
Instrument"
on
page 118
7. Restart your system by following the instructions in
"Shutting Down the TSQ 8000 Evo
Ion Guide Wires
"Replacing the Analyzer Tray in the TSQ 8000 Evo
to replace the tray into the instrument.
3
Advanced Troubleshooting
Replacing Components of the Analyzer
on
page
"Restarting the System"
on
TSQ 8000 Evo Hardware Manual
109.
Nut
Screw
page
120.
131