4
Installation for 990 Micro GC
Before you begin
Before you begin
Compressed liquid/gas sample cylinder safety
The Gasifier assists in the handling of high pressure liquid and gas samples. Make sure you are
familiar with the safe use of compressed sample cylinders.
Check ventilation condition
For safety and environmental protection, make sure the waste coming out of the gasifier and GC
vent connects to the ventilation system.
Tools and parts required for installation
•
Torx T-20 screwdriver, 1pcs
•
5/16 inch wrench, 1pcs
•
Filter Kit, P/N G3535-60008, 1pcs
•
Gasifier mounting bracket, P/N G3535-00016, 1 pcs
•
Screws M4*8, P/N 0515-2113, 3 pcs
•
USB Cable, P/N 8121-3060, 1 pcs
•
Core Ferrite, P/N 9170-2626, 1 pcs
Check compatibility to the GC system
Before connecting the gasifier, check that your system meets the following software
requirements.
Firmware and software
The gasifier requires micro GC firmware version 1.04 or greater.
The driver will support connection to/communication with the following platforms:
•
OpenLab CDS 2.1-2.5 (VL)
•
OpenLab ChemStation C.01.06-C.01.10 (VL)
•
OpenLab CDS EZChrom A.04.06-A.04.09 (VL&COMPACT)
Leak test
To start up the Gasifier for the first time, or after a long idle period, Agilent advises you do a leak
check for safety purposes. See
a detailed procedure.
30
Checking for Leaks
on page 53 of the
Agilent 8860, 8890, and 990 Micro GC Gasifier User Manual
Troubleshooting
section for