Maintaining the FID

To Perform Maintenance on the FID Collector Assembly

To Perform Maintenance on the FID Collector
Assembly
Perform only the steps and gather only the parts that apply to the desired maintenance
NOTE
task(s).
1 Gather the following:
To avoid contaminating the FID, wear clean, lint-free gloves when handling the collector
C AU T I O N
assembly.
2 Launch the GC maintenance wizard: Maintenance > Detectors, select the detector, then
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
WAR N IN G
hot, wear gloves to protect your hands.
3 Remove the FID ignitor. If you are not replacing the ignitor, skip to
150
Replacement ignitor assembly, see
Replacement ignitor castle
Two collector insulators
Collector
Spring washer
Gasket
T-20 Torx screwdriver
1/4-inch nut driver
Tweezers
5/16-inch wrench
Lint-free gloves
Clean cloth
select Perform Maintenance > Maintain FID Collector > Start Maintenance. The wizard will
guide you through the maintenance steps described below.
a Disconnect the ignitor cable assembly.
"Consumables and Parts for the FID"
on page 140.
step
5.
Maintaining Your GC