Quick Setup Procedure

N O TE
Agilent 86038A Optical Dispersion Analyzer, Third Edition

Quick Setup Procedure

For your convenience, a quick setup feature has been included on the ODA.
The Quick Setup key accesses a series of control panels that help you
specify the basic settings needed for making a measurement. A
measurement time estimation is provided so that you can trade off
measurement time versus the measurement settings.
The following procedure outlines the steps to make a precisely defined
measurement.
For experienced users, a detailed process can be used which provides
more flexibility than the Quick Setup procedure.
Measurement Process" on page
To get the most performance out of your ODA, it is recommended that you
read
Chapter 2, "Measurement
Measurements", and
Chapter 4, "Function
chapters cover the linkages between the different measurement that can
be made, specific devices being measured, and the setup parameters used
to make the measurements.
Setup the Normalization
• Connect the patch cord for the normalization.
• Check the connections with the Real Time Port Monitor.
Port Monitor" on page
Quick Setup Process
• On the front panel, press Quick Setup.
• Select DUT configuration and input ports.
• Choose which attributes to measure. Select stepped or swept mode.
• Set the measurement wavelength range.
• Set the resolution slider (lowest to highest resolution).
• Begin the Normalization
Measurement Process
• Connect the DUT.
• Begin the Measurement.
• View the data and manipulate the display and markers.
• Smooth or curve fit the data as desired.
• Save the measurement and/or export the data.