Procedure 1.

Vertical Alignment
Adjustment
NOTE
NOTE
Figure 3-1
This procedure applies to a CRT based display only.
The vertical alignment can be affected by magnetic interference. Before
adjusting the vertical position, be sure the analyzer is in a non-magnetic
environment and the CRT is degaussed.
Perform the following steps to adjust the vertical alignment.
1. Switch on the system and allow it to warm up for 60 minutes.
2. Remove the side panel nearest the display to access the vertical position
and focus controls. See
3. Insert a narrow, non-conductive, flathead screwdriver (2 or more inches
long) into the vertical position hole.
4. Adjust the control until the softkey labels align with the softkeys.
Vertical adjustment location
Making 8510XF Adjustments
Figure
3-1.
8510XF Service Quick Reference Guide 3-3