Marker X Axis Value (Remote Command Only); Marker X Axis Position (Remote Command Only) - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Initial S/W Revision

Marker X Axis Value (Remote Command Only)

Sets the marker X Axis value in the current marker X Axis Scale unit. It has no effect if the control mode
is
but is the SCPI equivalent of entering an X value if the control mode is
Off,
Mode
Remote Command
Example
Notes
Preset
State Saved
Range
Initial S/W Revision

Marker X Axis Position (Remote Command Only)

Sets the marker X position in trace points. It has no effect if the control mode is
equivalent of entering a value if the control mode is
X Axis Scale units. The entered value is immediately translated into the current X Axis Scale units for
setting the value of the marker.
Mode
Remote Command
Prior to A.02.00
BASIC
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
X <freq>
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
X?
CALC:SPEC:MARK3:X 100
CALC:SPEC:MARK3:X?
If no suffix is sent it uses the fundamental units for the current marker X Axis
Scale. If a suffix is sent that does not match the current marker X Axis Scale
unit, an error "Invalid suffix" is generated.
The query returns the marker's absolute X Axis value if the control mode is
Normal
, or the offset from the marker's reference marker if the control mode
is
Delta
. The query is returned in the fundamental units for the current marker
X Axis scale: Hz for
Frequency
. If the marker is
Time
Off
You must be in the IQ Analyzer (Basic) mode to use this command. Use
INSTrument:SELect to set the mode.
After a preset, all markers are turned OFF, so Marker X Axis Value query
returns a not a number (NAN).
No
Depends on X axis range of selected Trace.
Prior to A.02.00
Normal
BASIC
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
X:POSition <freq>
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
X:POSition?
Complex Spectrum Measurement
Normal
and
, seconds for
Inverse Time
the response is not a number.
Off,
or
Delta
- except in trace points rather than
Marker
or
.
Delta
and
Period
but is the SCPI
231

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9030aN9020aN9010aN9000aN9038a

Table of Contents