Compatible SCPI Commands
Supported SCPI Commands
Min
Max
SENSe:ROSCillator Subsystem
The SENSe:ROSCillator subsystem controls the reference oscillator. The numeric suffix in SENSe is
irrelevant for the commands in this subsystem.
[SENSe<1|2>:]ROSCillator:SOURce INTernal | EXTernal
Switches between an external or internal reference oscillator for frequency processing of external
generators 1 and 2.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Range
SENSe:SWEep Subsystem
The SENSe:SWEep subsystem controls the sweep parameters.
[SENSe<1|2>:]SWEep:TIME 2.5ms to 16000s (frequency domain) | 1us to 16000s (time domain)
Specifies sweep time.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Min
Max
148
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
1
3
[:SENSe]:ROSCillator:SOURce INTernal|EXTernal
[:SENSe]:ROSCillator:SOURce?
ROSC:SOUR EXT
FSE, FSP, FSU
INT
Yes.
INTernal|EXTernal
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
SWE:TIME 500ms
FSE, FSP, FSU
1ms
Saved in instrument state.
1 ms
4000 s