Status Reporting System

Turning Off the Status Reporting System

Turning Off the Status Reporting System
The E4981A status reporting system can be disabled. In a disabled condition,
the data in status reporting system is not updated. The benefit of disabling
Status Reporting System is that it shortens the End of Measurement (EOM)
time. See measurement time in Specification and Supplemental Performance
Characteristics in the E4981A User Guide for more details.
The Status Reporting System can be turned off using the
on page
271.
:STATus:OPERation:UPDate
372