Trigger[:Seq1]:Source - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

:TRIGger[:SEQ1]:SOURce

Syntax
:TRIGger[:SEQ1]:SOURce {INTernal|MANual|EXTernal|BUS}
:TRIGger[:SEQ1]:SOURce?
Description
Selects the trigger mode from the following four types.
Internal
Manual
External
BUS
Parameters
INTernal (initial value)
MANual
EXTernal
BUS
Query response
{INT|MAN|EXT|BUS}<newline><^END>
Related commands
*TRG
:READ?
Equivalent key
[Meas Setup] - TRIG - INT|MAN|EXT|BUS
sequence
You cannot set the trigger mode to GPIB/LAN/USB (Bus) using the front
panel keys.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Configures the instrument to use its internal trigger source so
that is it automatically and continuously triggered.
Configures the instrument to be triggered when you press the
key on the front panel.
[Trig]
Configures the instrument to be triggered when a trigger
signal is inputted through the Ext TRIG terminal or
handler/scanner interface.
Configures the instrument to be triggered when the
command is executed through GPIB/LAN/USB.
Description
Specifies Internal.
Specifies Manual.
Specifies External.
Specifies GPIB/USB/LAN (Bus).
on page 163
on page 230
*TRG
305