The Keysight E4981A provides error messages to indicate its operating status.
This appendix describes the error messages of the E4981A in order of error
number. To search for error messages alphabetically, refer to the Operation
Manual.

Error Messages (Order of Error Number)

Error messages are displayed in the lower row of the E4981A's display. You can
read them out by using the SCPI command. This section provides a description
of each error message and its remedy.
Errors with a negative error number are basically general errors defined by
IEEE488.2 for GPIB/LAN/USB instruments. On the other hand, errors with a
positive error number are defined specifically for the E4981A.
16
Reference Measurement Aborted
This error occurs when REF data measurement is aborted.
21
1 MHz opt. not installed
This error occurs when cable correction commands are sent to E4981A with
Option 002, via GPIB/LAN/USB. This error is not generated with front panel
operation.
41
Correction Measurement Aborted
This error occurs when the correction data measurement is aborted.
43
Measurement failed
A measurement failure has occurred during measuring the correction data.
46
LOAD measurement incomplete
This error occurs when the cable correction LOAD measurement is incomplete.
47
OPEN measurement incomplete
This error occurs when the cable correction OPEN measurement is incomplete.