Command Reference
E4981A SCPI Commands
[:SENSe][:FIMPedance]:RANGe:AUTO
Syntax
[:SENSe][:FIMPedance]:RANGe:AUTO {ON|OFF|1|0}
[:SENSe][:FIMPedance]:RANGe:AUTO?
Description
Selects the measurement range mode from the auto range (automatic range
switching) or hold range (fixed range).
Setting the measurement range (set with the
[:SENSe][:FIMPedance]:RANGe[:UPPer]
the hold range mode.
Parameters
ON or 1 (initial value)
OFF or 0
Query response
{1|0}<newline><^END>
Related commands
[:SENSe][:FIMPedance]:RANGe[:UPPer]
Equivalent key
[Meas Setup] - RANGE - AUTO|HOLD
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command) automatically selects
Description
Specifies the auto range mode.
Specifies the hold range mode.
on page 266