Syntax
:CALCulate1:FORMat {CP|CS}
:CALCulate1:FORMat?
Description
Specifies the primary parameter to be measured.
When the secondary parameter has been set to G or RP and the primary
parameter is set to CS, the secondary parameter is automatically changed to
D. Also, when the secondary parameter has been set to RS and the primary
parameter is set to CP, the secondary parameter is automatically changed to
D.
Parameters
CP (initial value)
CS
Query response
{CP|CS}<newline><^END>
Related commands
:CALCulate2:FORMat
Equivalent key
[Meas Setup]- FUNC
sequence
188
Description
Specifies the capacitance value measured using the parallel
equivalent circuit model for the primary parameter.
Specifies the capacitance value measured using the series
equivalent circuit model for the primary parameter.
on page 191