Trigger[:Seq1][:Immediate]; Trigger[:Seq1]:Slope - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

:TRIGger[:SEQ1][:IMMediate]

Syntax
:TRIGger[:SEQ1][:IMMediate]
Description
Immediately generates a trigger and executes a measurement. This command
is valid for trigger source values of MANual or BUS trigger specified by
:TRIGger[:SEQ1]:SOURce
INTernal or EXTernal, this command generates an error.
If the trigger system is not in the trigger wait state (trigger event detection
state), executing this command causes an error and the command is ignored.
For details on the trigger system, refer to
query)
Equivalent key
[Trigger]
sequence

:TRIGger[:SEQ1]:SLOPe

Syntax
:TRIGger[:SEQ1]:SLOPe {POSitive|NEGative}
:TRIGger[:SEQ1]:SLOPe?
Description
Determines the trigger polarity for BNC external trigger on the rear panel. This
does not affects the trigger signal on handler and scanner interface.
Related commands
*TRG
Parameters
POSitive (Initial [Factory
set] value)
NEGative
The value of this command is not changed with
and is only changed with Factory Default reset (available only through
Front Panel).
{POS|NEG}<newline><^END>
Query response
Equivalent key
[System] - EXT TRIG POL - POS|NEG
sequence
304
on page 163
Description
Detects positive edge trigger signal.
Detects negative edge trigger signal.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command. On setting the trigger source value to
"Trigger system" on page
*RST
57. (No
and
:SYSTem:PRESet

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a/001E4981a/002

Table of Contents