Sorting Based on Measured Result (Comparator Function)
Setting Up Comparator Function
Setting the range's lower and upper limit values
To set the lower limit value and upper limit value of the primary parameter limit
range (BIN1 to BIN9), use the following command.
To set the lower limit value and upper limit value of the secondary parameter
limit range, use the following command.
If you set the upper limit value to a value equal to or less than the lower
limit value, no limit range is used. Doing this would be equivalent to setting
the limit range to OFF.
As shown in
judgment is performed in increasing order of BIN number. Therefore, when
the limit ranges of two BINs overlap, the result is sorted to the BIN with the
smaller number.
For the tolerance mode, the reference value does not need to be within the
limit range (between the lower limit value and upper limit value).
Gaps are allowed between the limit ranges.
When the low limit value is greater than the upper limit, a warning
message "improper high/low limits" is d isplayed.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
:CALCulate1:COMParator:PRIMary:BIN{1-9}[:LIMIT]
:CALCulate1:COMParator:SECondary:LIMit
Figure 7-1, "Sorting judgment flow," on page
on page 182
on page 185
79, the sorting
75