Turn-on and Software Installation
Turning the System Off
To shut down the test system, turn the power switch to the left (so that it is
horizontal).
Emergency
Stop
Reset
Button.
Press after
turning
Power
switch
Figure 2
Light Tower Inside the Safety Hood
Green Light -- High voltage is off. It is safe to open the Safety Hood
Amber Light -- Emergency Shut-down button has been pushed. Pull the
Red Light -- Safety Hood is locked and cannot be opened. High Voltage may be
Emergency Off Functionality
The Emergency Stop button functionality works in two steps:
- When pressed, it interrupts the interlock of the high voltage source and
- To reset, pull out to release the Emergency Off button. Reset the Emergency Off
50
On
.
Rack and Safety Enclosure Front Panel Controls
(Power switch shown in the On position)
Emergency Off button out and press the RESET button.
present.
unlocks the safety switch. If the Safety Hood is closed, It keeps the enclosure
locked. After about 1 second the high voltage source shuts off by interrupting
the main power line.
circuit by pressing the Quit button.
Step 1: Connect LAN and Power Cable to Rack
Keysight Double-Pulse Test System Installation Guide
Power Switch
(shown in the
On position)
Open Hood.
Press button
while opening
the safety
hood.