:CALCulate1:COMParator[:STATe]

Syntax
:CALCulate1:COMParator[:STATe] {ON|OFF|1|0}
:CALCulate1:COMParator[:STATe]?
Description
Turns ON/OFF the comparator function.
This setup is interlocked with the ON/OFF state of the signal output of the
handler interface.
Changing the measurement parameter will automatically turn off this
function. To avoid this, you have to write a program so that this command
is executed after the execution of the measurement parameter setup
command
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:FORMat
:CALCulate2:FORMat
Equivalent key
[Meas Setup] - LIMIT TABLE - COMP
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
(:CALCulate1:FORMat
Description
Turns ON the comparator function.
Turns OFF the comparator function.
on page 188
on page 191
and :CALCulate2:FORMat).
187