Preparation for Accurate Measurement (Executing
Correction)
This chapter describes how to execute the correction function.

Executing OPEN/SHORT/LOAD Correction

Turning ON/OFF correction function

To turn ON/OFF each type of correction, use the corresponding command
below.
Type of correction
OPEN correction
SHORT correction
LOAD correction
Executing the measurement of correction data with the
[:SENSe]:CORRection:COLLect[:ACQuire]
automatically measures data and turns ON the correction.
The measured value and the standard value is available to display or input
using the following parameters:
Type of correction
OPEN correction
SHORT correction
LOAD correction
Command
[:SENSe]:CORRection:OPEN[:STATe]
page 248
[:SENSe]:CORRection:SHORt[:STATe]
page 249
[:SENSe]:CORRection:LOAD[:STATe]
page 242
Parameter
G-B, Cp-G
R-X, Ls-Rs
Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs
on
on
on
command on page 237