[:Sense][:Fimpedance]:Contact1:Verify:threshold1 - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1

Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 <numeric>
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1?
Description
Sets/Gets the Threshold1 value for Contact check. Threshold1 is a parameter
which is proportional contact resistance of Hp or Hc. It is recommended to use
the default (initial) value.
The available range for Hp, Hc and Lc is 220 µF to 1 mF at 120 Hz and
22 µF to 100 µF at 1 kHz. The contact resistance for Lp cannot be detected
by contact check feature for this range.
Parameters
Description
Range
Initial value
Resolution
Query response
<numeric><newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2
Equivalent key
[Meas Setup] - CONT CHK- CC1 TH1
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
<numeric>
Threshold1 value for Contact1.
0 to 1
0.1
0.01
on page 260
on page 262
261

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a/001E4981a/002

Table of Contents