[:Sense]:Correction:collect[:Acquire] - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:COLLect[:ACQuire]

Syntax
[:SENSe]:CORRection:COLLect[:ACQuire] {STAN1|STAN2|STAN3}
Description
Measures the correction data for OPEN/SHORT/LOAD correction and turns on
the correction function.
For Open and Short correction, the correction data is measured for all
measurement frequencies. For load correction, the correction data is measured
for the measurement frequency used in executing the command (set with the
:SOURce:FREQuency[:CW]
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
the correction data for the selected channel when executing the command
(selected with the
(No query)
Parameters
STAN1
STAN2
STAN3
Related commands
[:SENSe]:CORRection:OPEN[:STATe]
[:SENSe]:CORRection:SHORt[:STATe]
[:SENSe]:CORRection:LOAD[:STATe]
[:SENSe]:CORRection:MULTiple[:STATe]
[:SENSe]:CORRection:MULTiple:CHANnel
Equivalent key
[Meas Setup] - CORRECTION - OPEN - MEAS OPEN
sequence
[Meas Setup] - CORRECTION - SHORT - MEAS SHORT
[Meas Setup] - CORRECTION - LOAD - MEAS LOAD
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command).
[:SENSe]:CORRection:MULTiple:CHANnel
Description
Specifies the OPEN correction.
Specifies the SHORT correction.
Specifies the LOAD correction.
command), this is measured as
on page 248
on page 249
on page 242
on page 245
on page 243
command).
237