[:Sense]:Correction:offset[:State] - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:OFFSet[:STATe]

Syntax
[:SENSe]:CORRection:OFFSet[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:OFFSet[:STATe]?
Description
Turns ON/OFF the offset correction function.
Changing any of the measurement parameter will automatically turn off
this function. To avoid this, you have to write a program so that this
command is executed after the execution of the measurement parameter
setup command
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands
[:SENSe]:CORRection:OFFSet:DATA
:CALCulate1:FORMat
:CALCulate2:FORMat
Equivalent key
[Meas Setup] - CORRECTION - OFFSET -ON|OFF
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
(:CALCulate1:FORMat
Description
Turns ON the offset correction function.
Turns OFF the offset correction function.
on page 246
on page 188
on page 191
and :CALCulate2:FORMat).