D
Operations when Overload, No Contact, or Low C
is Detected
This appendix describes display output, GPIB/LAN/USB output, and handler
interface output when an Overload, No contact, or Low C is detected.

Operations when Overload/No Contact/Low C is Detected

Table D-5
detected:
shows operations of the E4981A when one of the following items is
Overload
When the available measurement range is exceeded more than about
10% (refer to "Specifications and Supplemental Performance
Characteristics" in Operation Manual)
When connection between the UNKNOWN terminals to DUT is not
correct
When SLC is ON, if the applied signal level for DUT is dropped more than
10% from the setting level
No contact:
When the contact check parameter is over the specified threshold
Low C:
When measured primary parameter result is equal to or less than the
boundary value specified for the Low C reject function
Out of display range:
When measured result exceeds the allowable display range (refer to
"Specifications and Supplemental Performance Characteristics" in
Operation Manual) (regardless of fixed or floating point display)
When measured result exceeds the allowable display range for the fixed
point display